CVE-2025-15653CWE-668

Dräger Zeus IE Anesthesia Workstation USB Interface Privilege Escalation

High · published June 2, 2026

CVSS v4.0
7.0
EPSS
0%
Percentile
6.0
In the wild
Unconfirmed
What it is

Dräger Zeus Infinity Empowered (Zeus IE) and Zeus RS C500 anesthesia workstations contain a local security vulnerability that allows unauthorized individuals with physical access to compromise software integrity via USB interface manipulation. Attackers can exploit the unprotected USB interfaces to impair therapy functions, manipulate device-processed data, or leverage the device as a pivot point for broader network-based attacks when connected to a network or Dräger Service Connect.

The record
Technical detail
CVSS v4.0
7.0 · HIGH
Vector
CVSS:4.0/AV:P/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N
EPSS
0.00169 · 6.0th percentile
Weakness
CWE-668 · Exposure of Resource to Wrong Sphere