CVE-2025-0683CWE-359

Exposure of Private Personal Information to an Unauthorized Actor vulnerability in Contec Health CMS8000 Patient Monitor

High · published January 30, 2025

CVSS v4.0
8.2
EPSS
1%
Percentile
55.9
In the wild
Unconfirmed
What it is

In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text

patient data to a hard-coded public IP address when a patient is hooked

up to the monitor. This could lead to a leakage of confidential patient

data to any device with that IP address or an attacker in a

machine-in-the-middle scenario.

The record
Technical detail
CVSS v4.0
8.2 · HIGH
Vector
CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:N/VA:N/SC:N/SI:N/SA:N
EPSS
0.00792 · 55.9th percentile
Weakness
CWE-359 · Exposure of Private Personal Information to an Unauthorized Actor
Published
2025-01-30T18:17Z
EPSS history
Timeline
  • 30 JAN 18:17Z
    Exposure of Private Personal Information to an Unauthorized Actor vulnerability in Contec Health CMS8000 Patient Monitor
    cvelistv5