CVE-2020-25187CWE-122

Medtronic MyCareLink Smart Heap-based Buffer Overflow

High · published December 14, 2020

CVSS v3.1
8.8
EPSS
4%
Percentile
90.3
In the wild
Unconfirmed
What it is

Medtronic MyCareLink Smart 25000 is

vulnerable when an authenticated attacker runs a debug command, which can be sent to the patient reader and cause a heap overflow event within the MCL Smart Patient Reader software stack. The heap overflow could allow an attacker to remotely execute code on the MCL Smart Patient Reader, potentially leading to control of the device

The record
Technical detail
CVSS v3.1
8.8 · HIGH
Vector
CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:C/C:H/I:H/A:H
CVSS v4.0
Not supplied
EPSS
0.03931 · 90.3th percentile
Weakness
CWE-122 · Heap-based Buffer Overflow
Published
2020-12-14T19:18Z
EPSS history
Timeline
  • 14 DEC 19:18Z
    Medtronic MyCareLink Smart Heap-based Buffer Overflow
    cvelistv5