CVE-2020-25183CWE-287

Medtronic MyCareLink Smart Improper Authentication

High · published December 14, 2020

CVSS v3.1
8.0
EPSS
1%
Percentile
56.0
In the wild
Unconfirmed
What it is

Medtronic MyCareLink Smart 25000 contains

an authentication protocol vulnerability where the method used to authenticate between the MCL Smart Patient Reader and the Medtronic MyCareLink Smart mobile app is vulnerable to bypass. This vulnerability enables an attacker to use another mobile device or malicious application on the patient’s smartphone to authenticate to the patient’s Medtronic Smart Reader, fooling the device into believing it is communicating with the original Medtronic smart phone application when executed within range of Bluetooth communication.

The record
Technical detail
CVSS v3.1
8.0 · HIGH
Vector
CVSS:3.1/AV:A/AC:L/PR:N/UI:R/S:U/C:H/I:H/A:H
CVSS v4.0
Not supplied
EPSS
0.00796 · 56.0th percentile
Weakness
CWE-287 · Improper Authentication
Published
2020-12-14T19:18Z
EPSS history
Timeline
  • 14 DEC 19:18Z
    Medtronic MyCareLink Smart Improper Authentication
    cvelistv5