CVE-2019-5024CWE-693

A restricted environment escape vulnerability exists in the “kiosk mode” function of Capsule Technologies SmartLinx Neuron 2 medical information collection…

High · published April 11, 2019

CVSS v3.0
7.6
EPSS
0%
Percentile
39.9
In the wild
Unconfirmed
What it is

A restricted environment escape vulnerability exists in the “kiosk mode” function of Capsule Technologies SmartLinx Neuron 2 medical information collection devices running versions 9.0.3 or lower. A specific series of keyboard inputs can escape the restricted environment, resulting in full administrator access to the underlying operating system. An attacker can connect to the device via USB port with a keyboard or other HID device to trigger this vulnerability.

The record
Technical detail
CVSS v3.0
7.6 · HIGH
Vector
CVSS:3.0/AV:P/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
CVSS v4.0
Not supplied
EPSS
0.00472 · 39.9th percentile
Weakness
CWE-693 · Protection Mechanism Failure
Published
2019-04-11T17:45Z
EPSS history
Timeline
  • 11 APR 17:45Z
    A restricted environment escape vulnerability exists in the “kiosk mode” function of Capsule Technologies SmartLinx Neuron 2 medical information collection…