CVE-2019-25724CWE-400

Dräger Infinity M300 VG2.x Network-Based Denial of Service

High · published June 2, 2026

CVSS v4.0
7.1
EPSS
0%
Percentile
6.6
In the wild
Unconfirmed
What it is

Dräger Infinity M300 patient worn monitors with software version VG2.x and earlier contain a network-based denial of service vulnerability that allows attackers with access to the hospital or Infinity Network to repeatedly trigger device reboots until the device enters a fail state requiring manual restart. Attackers can exploit this vulnerability to cause loss of wireless network connectivity, temporary loss of patient monitoring, and interruption of alarm functionality until the device is manually recovered.

The record
Technical detail
CVSS v4.0
7.1 · HIGH
Vector
CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:N/VI:N/VA:H/SC:N/SI:N/SA:N
EPSS
0.00175 · 6.6th percentile
Weakness
CWE-400 · Uncontrolled Resource Consumption
Published
2026-06-02T19:11Z
EPSS history
Timeline
  • 02 JUN 19:11Z
    Dräger Infinity M300 VG2.x Network-Based Denial of Service
    cvelistv5