CVE-2019-25722CWE-798

Dräger SC Monitoring Devices Hard-coded Credentials and DoS

High · published June 2, 2026

CVSS v4.0
7.2
EPSS
0%
Percentile
8.5
In the wild
Unconfirmed
What it is

Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain hard-coded plaintext credentials in source code and a denial-of-service vulnerability that allows local and remote attackers to compromise device integrity across all software versions. A local attacker with direct device access can use the hard-coded credentials to access service and clinical accounts and alter device configuration, while a remote attacker can send malformed network packets to cause repeated device reboots, ultimately resulting in loss of network connectivity and disruption of patient monitoring.

The record
Technical detail
CVSS v4.0
7.2 · HIGH
Vector
CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N
EPSS
0.00193 · 8.5th percentile
Weakness
CWE-798 · Use of Hard-coded Credentials