CVE-2019-10964CWE-284

Medtronic MiniMed 508 and Paradigm Series Insulin Pumps Improper Access Control

High · published June 28, 2019

CVSS v3.1
7.1
EPSS
1%
Percentile
68.8
In the wild
Unconfirmed
What it is

Medtronic MiniMed Insulin Pumps

are designed to communicate using a wireless RF with other devices, such as blood glucose meters, glucose sensor transmitters, and CareLink USB devices. This wireless RF communication protocol does not properly implement authentication or authorization. An attacker with adjacent access to one of the affected insulin pump models can inject, replay, modify, and/or intercept data. This vulnerability could also allow attackers to change pump settings and control insulin delivery.

The record
Technical detail
CVSS v3.1
7.1 · HIGH
Vector
CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:H/A:H
CVSS v4.0
Not supplied
EPSS
0.01234 · 68.8th percentile
Weakness
CWE-284 · Improper Access Control
Published
2019-06-28T20:58Z