CVE-2018-8870CWE-259
Medtronic MyCareLink Patient Monitor Use of Hard-coded Password
Medium · published July 2, 2018
What it is
Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains a hard-coded operating system password. An attacker with physical access can remove the case of the device, connect to the debug port, and use the password to gain privileged access to the operating system.
The record
Technical detail
- CVSS v3.1
- 6.4 · MEDIUM
- Vector
- CVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H
- CVSS v4.0
- Not supplied
- EPSS
- 0.00362 · 28.8th percentile
- Weakness
- CWE-259 · Use of Hard-coded Password
- Published
- 2018-07-02T18:00Z
EPSS history
Timeline