CVE-2018-8868CWE-749

Medtronic MyCareLink Patient Monitor Exposed Dangerous Method or Function

Medium · published July 2, 2018

CVSS v3.1
6.2
EPSS
0%
Percentile
25.5
In the wild
Unconfirmed
What it is

Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can exploit other vulnerabilities to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality.

The record
Technical detail
CVSS v3.1
6.2 · MEDIUM
Vector
CVSS:3.1/AV:P/AC:H/PR:L/UI:N/S:C/C:H/I:L/A:L
CVSS v4.0
Not supplied
EPSS
0.00333 · 25.5th percentile
Weakness
CWE-749 · Exposed Dangerous Method or Function
Published
2018-07-02T18:00Z
EPSS history
Timeline
  • 02 JUL 18:00Z
    Medtronic MyCareLink Patient Monitor Exposed Dangerous Method or Function
    cvelistv5