CVE-2018-14781CWE-294

Medtronic MiniMed MMT-500/MMT-503 Remote Controllers Authentication Bypass by Capture-replay

Medium · published August 13, 2018

CVSS v3.1
5.3
EPSS
1%
Percentile
53.1
In the wild
Unconfirmed
What it is

Medtronic MiniMed MMT

devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.

The record
Technical detail
CVSS v3.1
5.3 · MEDIUM
Vector
CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:N
CVSS v4.0
Not supplied
EPSS
0.00712 · 53.1th percentile
Weakness
CWE-294 · Authentication Bypass by Capture-replay
Published
2018-08-13T22:00Z
EPSS history
Timeline
  • 13 AUG 22:00Z
    Medtronic MiniMed MMT-500/MMT-503 Remote Controllers Authentication Bypass by Capture-replay
    cvelistv5