CVE-2018-14781CWE-294
Medtronic MiniMed MMT-500/MMT-503 Remote Controllers Authentication Bypass by Capture-replay
Medium · published August 13, 2018
What it is
Medtronic MiniMed MMT
devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.
The record
Technical detail
- CVSS v3.1
- 5.3 · MEDIUM
- Vector
- CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:N
- CVSS v4.0
- Not supplied
- EPSS
- 0.00712 · 53.1th percentile
- Weakness
- CWE-294 · Authentication Bypass by Capture-replay
- Published
- 2018-08-13T22:00Z
EPSS history
Timeline