CVE-2018-10596CWE-923

Medtronic 2090 Carelink Programmer Improper Restriction of Communication Channel to Intended Endpoints

High · published July 2, 2018

CVSS v3.1
7.1
EPSS
1%
Percentile
70.4
In the wild
Unconfirmed
What it is

Medtronic 2090 CareLink Programmer

uses a virtual private network connection to securely download updates. It does not verify it is still connected to this virtual private network before downloading updates. The affected products initially establish an encapsulated IP-based VPN connection to a Medtronic-hosted update network. Once the VPN is established, it makes a request to a HTTP (non-TLS) server across the VPN for updates, which responds and provides any available updates. The programmer-side (client) service responsible for this HTTP request does not check to ensure it is still connected to the VPN before making the HTTP request. Thus, an attacker could cause the VPN connection to terminate (through various methods and attack points) and intercept the HTTP request, responding with malicious updates via a man-in-the-middle attack. The affected products do not verify the origin or integrity of these updates, as it insufficiently relied on the security of the VPN. An attacker with remote network access to the programmer could influence these communications.

The record
Technical detail
CVSS v3.1
7.1 · HIGH
Vector
CVSS:3.1/AV:N/AC:H/PR:L/UI:R/S:U/C:H/I:H/A:H
CVSS v4.0
Not supplied
EPSS
0.01320 · 70.4th percentile
Weakness
CWE-923 · Improper Restriction of Communication Channel to Intended Endpoints
Published
2018-07-02T18:00Z
EPSS history
Timeline
  • 02 JUL 18:00Z
    Medtronic 2090 Carelink Programmer Improper Restriction of Communication Channel to Intended Endpoints
    cvelistv5