CVE-2017-12712CWE-287

The authentication algorithm in Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017, which involves an authentication key and time stamp, can be…

published April 25, 2018

CVSS
—
EPSS
1%
Percentile
64.0
In the wild
Unconfirmed
What it is

The authentication algorithm in Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017, which involves an authentication key and time stamp, can be compromised or bypassed, which may allow a nearby attacker to issue unauthorized commands to the pacemaker via RF communications. CVSS v3 base score: 7.5, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.

The record
Technical detail
CVSS
Not scored
CVSS v4.0
Not supplied
EPSS
0.01049 · 64.0th percentile
Weakness
CWE-287 · Improper Authentication
Published
2018-04-25T13:00Z
EPSS history
Timeline
  • 25 APR 13:00Z
    The authentication algorithm in Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017, which involves an authentication key and time stamp, can be…