CVE-2014-5431CWE-259

Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 contains a hard-coded password, which…

published March 26, 2019

CVSS
—
EPSS
0%
Percentile
30.5
In the wild
Unconfirmed
What it is

Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 contains a hard-coded password, which provides access to basic biomedical information, limited device settings, and network configuration of the WBM, if connected. The hard-coded password may allow an attacker with physical access to the device to access management functions to make unauthorized configuration changes to biomedical settings such as turn on and off wireless connections and the phase-complete audible alarm that indicates the end of an infusion phase. Baxter has released a new version of the SIGMA Spectrum Infusion System, version 8, which incorporates hardware and software changes.

The record
Technical detail
CVSS
Not scored
CVSS v4.0
Not supplied
EPSS
0.00378 · 30.5th percentile
Weakness
CWE-259 · Use of Hard-coded Password
Published
2019-03-26T15:37Z
EPSS history
Timeline
  • 26 MAR 15:37Z
    Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 contains a hard-coded password, which…
    cvelistv5